B. DC or AC Current Measurement
(See figure 4)
Warning
Before connecting the meter in series with the circuit
under test, be sure power to the circuit is off to avoid
the possibility of arcing.
If the fuse burns out during measurement, the Meter
may be damaged and there is risk of personal injury
to the operator.
Use proper terminals, function, and range for the
measurement. When the test leads are connected to
the current terminals, do not parallel them across any
circuit.
To measure current, do the following:
18
Model 72-1016: OPERATING MANUAL
1.
2.
3.
4.
Insert the red test lead into the
µA mA
terminal and
the black test lead into the
COM
terminal.
Set the rotary switch to an appropriate measurement
position in
µA
,
mA
or
A
, press
SELECT
button
to select AC or DC measurement mode.
Connect the test leads in series with the object being
measured. The measured value shows on the display.
AC measurement displays True RMS value.
Press
AC/AC+DC
button to measure AC+DC
current’s true RMS
(Figure 4)