21
Note
When measuring low resistances, the test leads and
internal wiring will add approximately 0.2 ~ 0.5
of error. To obtain accurate readings in low-resistance,
short-circuit the test lead beforehand and record the
reading obtained, call this reading as X. Then use
the equation:
measured resistance value (Y) – (X) = accurate
readings of resistance.
If reading with shorted test leads is not < 0.5 ,
check for loose test leads or possible incorrect
function selection.
When measuring high resistance (>1M ), it is normal
to take several seconds to obtain a stable reading.
In order to obtain stable reading, choose shorter test
lead to carrying out measurement.
The LCD displays
OL
indicating open-circuit for the
tested resistor or the resistor value is higher than the
maximum range of the Meter.
Model 72-1016: OPERATING MANUAL
When resistance measurement has been completed,
disconnect the connection between the test leads and
the circuit under test, and remove the test leads from
the input terminals if the meter.