46
T
able
6. Guidance and Manufacturer’s Declaration, Electromagnetic Immunity, for the
Tenex Health TX System.
The Tenex Health TX System is intended for use in the electromagnetic environment specified below. The customer
or the user of the Tenex Health TX System should assure that it is used in such an environment.
Immunity test
EN 60601 test level
Compliance level
Electromagnetic
Environment: guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic materials, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
±1 kV for input/output lines
± 2 kV for power supply
lines
±1 kV for input/output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
2 kV line to ground
1 kV line to line
2 kV line to ground
1 kV line to line
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines
IEC 61000-4-11
< 5% UT
(< 95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(< 95% dip in UT)
for 5 sec
< 5% UT
(< 95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(< 95% dip in UT)
for 5 sec
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the Tenex Health TX
System requires continued
operation during power
mains interruptions, it is
recommended that the
Tenex Health TX System
be powered from an
uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3A (rms)/m
3A (rms)/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE:
UT is the AC mains voltage prior to application of the test level.