![Tenex MicroTip TX1 Operator'S Manual Download Page 14](http://html1.mh-extra.com/html/tenex/microtip-tx1/microtip-tx1_operators-manual_1086479014.webp)
8
Manufacturer
TENEX HEALTH
Model
TX
Classification (IEC 60601)
CLASS I
PARAMETER
SPECIFICATION
Electrical
INPUT VOLTAGE
100 to 240 VAC @ 50/60 Hz
MAXIMUM CURRENT
1.2 Amps
LINE FUSES
5A T (Slow-blow)
MAX. OUTPUT VOLTAGE (CUTTING)
250 Volts
MAX. OUTPUT VOLTAGE (FOOT PEDAL)
5 Volts
Irrigation
FLUID DELIVERY
Pressure fed: regulated compressor
OPERATING PRESSURE
4 PSI nominal
VALVE TYPE
Solenoid-actuated Pinch Valve
CONTROL
Foot Pedal
Cutting
TX MicroTip TYPE
Sterile, Disposable Ultrasonic
OPERATING FREQUENCY
25–28 kHz
CONTROL
Foot Pedal
SYSTEM PRIMING
Automated Prime Cycle
Available Power Delivery
LOW
100% delivered at 50 ms ON and 50 ms OFF
MEDIUM
100% delivered at 75 ms ON and 25 ms OFF
HIGH
100% delivered Continuously
DUTY CYCLE (USER-CONTROLLED)
15s/45s (15 seconds ON, 45 seconds OFF at maximum power)
Aspiration
ASPIRATION PUMP
Peristaltic, Low Pulsation
AVAILABLE MAX. VACUUM LEVEL
User selectable: Low/Medium/High (100/300/500 mmHg)
AVAILABLE FLOW RATE
User selectable: Low/Medium/High (10/20/30 cc/min)
CONTROL
Foot Pedal
Dimensions (Console)
HEIGHT
6.5 inches / 16.5 centimeters
WIDTH
14.5 inches / 36.8 centimeters
DEPTH
13.5 inches / 34.3 centimeters
WEIGHT
21 lbs.
LCD TOUCH PANEL DISPLAY
12.1 inches diagonal / 280(H) × 218(V)
millimeters 800 x 600 pixels, color
d. TECHNICAL SPECIFICATIONS
Summary of Contents for MicroTip TX1
Page 1: ...TX Console Operator s Manual For Software Versions 0 1 4 and 1 0 0...
Page 2: ......
Page 6: ......
Page 55: ......