30
9. Ensure that the aspiration tube is properly centered on the bottom side of the cartridge and
then insert cartridge into side of console, pressing straight down firmly until locked in place.
NOTE: The TX Console is not sterile.
10. Align the Red Dot on the TX MicroTip electrical connector with the Red Dot on the console
side panel CUT receptacle, and then insert straight into the receptacle.
CAUTION:
Failure to align the Red Dot on the MicroTip connector with the Red Dot on the
console receptacle may damage the connector pins.
11. Spike saline irrigation bag.
12.
With the tip cap on, hold the MicroTip handpiece in a vertical position (needle pointing
upward) and press the PRIME key on the TX Console.
13. After successful priming, leave the tip cap on. Select IRRIGATION to ON, ASPIRATION
to MEDIUM or HIGH, CUTTING POWER to MEDIUM or HIGH and COAGULATION to
OFF, then depress foot pedal to test for acoustic signal.
CAUTION:
Ensure successful priming cycle and presence of acoustic signal prior to incising
the patient.
WARNING:
DO NOT check function of the MicroTip by placing hand or finger against the tip or
unintended damage to healthy tissue may result.
NOTE:
For a description of setup related error messages, consult the Troubleshooting and
Factory Default Settings sections of the TX Console Operator’s Manual.
F
igure
12
. CORRECT Aspiration Tube Alignment
F
igure
13
. CORRECT Insertion
F
igure
14
. INCORRECT Aspiration Tube Alignment
F
igure
15
. INCORRECT Insertion
F
igure
16
. Electrical Connector Orientation