CAUTION (REFER TO FIG 4-6 AND FIG 4-7):
The VIRALERT 3 system is
supplied with the Blackbody Reference Source installed so that it is visible
in the right hand portion of the images from the camera and thermal imager.
In the thermal image, the Blackbody Reference Source must be visible
within the gray Reference Area band on the right. It must not encroach into
the Measurement Area.
If your Blackbody Reference Source moves out of position, you will need to
adjust its mounting so that it is aligned correctly.
5)
Access to the
Settings
menu, in which you can adjust the temperature alarm
value and units of measurement, is password protected.
The default (factory-set) password is
0000
.
WARNING:
The threshold for an elevated body temperature alarm is a
critical parameter for use of VIRALERT 3.
You should consult a medical advisor for assistance in determining the
appropriate alarm point. The factory default value is 100.4 °F (38 °C).
For more information on the Settings menu and other features of the
software, refer to the Section 5 of this User Guide.