3 SYSTEM SETUP
3.1 Measurement Location
1)
Choose a location near the entrance of the area you want to protect.
2)
There must be space to install the Camera and Blackbody Reference Source
so that the Measurement Subject can face the Camera from a distance
of 3 ft/1m.
3)
Avoid locations where there are possible heat sources e.g. electrical
equipment, radiators, etc. in the field of view of the camera.
4)
Also avoid locations where the camera may pick up reflections from
windows, mirrors, etc.
5)
VIRALERT 3 is designed to scan one person at a time. The measurement
location should not be a high-traffic area where people other than the
subject will be in the field of view.
6)
During the temperature measurement process, the Subject will need to stand
still in front of the camera for approximately 2 seconds. Make sure there is
adequate provision for social distancing in your chosen measurement location.
7)
The camera must have a clear, unobstructed view of the Subject’s face,
ideally with the camera at a height around eye level. If necessary, use one of
the mounting accessories available from Telguard (see Section 1.3).
8)
The Blackbody Reference Source requires a power supply. You will also need
to provide power to the Laptop/PC running the VIRALERT 3 software.
9)
Once the system is installed and operating, use cable ties to secure the
cables to suitable structures. This minimizes the risk of the cables becoming
dislodged and reduces the risk of tripping over the cables.
10) Posters which explain the screening process and provide instructions for your
staff and visitors on how to use the system are available for download from:
www.Telguard.com/viralert3
3.2 System Connections
NOTE:
If you have purchased a mounting accessory i.e. Surface Mount Bracket, Clamp
Mount Bracket, or Tripod from Telguard, then each accessory will come supplied with
its own installation instructions.
Camera Connection
Connect the cylindrical connector on the
power and signal cable to the Camera.
3 SYSTEM SETUP