Getting Started Manual
922138-00 Rev A
37
B
excursion of the AC power line.
Touch inside the
Level
field. In the pop-up numeric
keypad, enter a value in millivolts or use the
up/down buttons to increase or decrease the value
in increments of 1 mV. Or move the slider bar to
increase or decrease the values. See Dialog Menu
section for more information on the slider bar.
Touch inside the
Coupling
field and select a coupling
type.
o
DC
– All the signal’s frequency components
are coupled to the trigger circuit for high
frequency bursts or where the use of AC
coupling would shift the effective trigger
level.
o
AC
– The signal is capacitively coupled. DC
levels are rejected, and frequencies below 50
Hz are attenuated.
o
LFREJ
- The signal is coupled through a
capacitive high-pass filter network, DC is
rejected and signal frequencies below 50 kHz
are attenuated. For stable triggering on
medium-to-high frequency signals.
o
HFREJ
- Signals are DC coupled to the trigger
circuit and a low-pass filter network
attenuates frequencies above 50 kHz. This is
used for triggering on low frequencies.
Choose Positive, Negative, or Window slope.
Window slope sets a threshold above and below the
trigger level beyond which the signal must pass to
generate a trigger. The slope can be either positive or
negative.
Summary of Contents for waverunner mxi-a series
Page 1: ...Getting Started Manual WaveRunner Xi A WaveRunner MXi A Series Oscilloscopes...
Page 2: ......
Page 3: ...WaveRunner Xi A MXi A Series Oscilloscopes Getting Started Manual January 2013...
Page 22: ...WaveRunner Xi A and MXi A Oscilloscopes 14 922138 00 Rev A Front Panel Controls...
Page 126: ...WaveRunner Xi A and MXi A Oscilloscopes 118 922138 00 Rev A...
Page 127: ......
Page 128: ......