Getting Started Manual
922138-00 Rev A
95
B
Acquisition
Offset Control
As you change the gain, this control allows you to either keep the vertical
offset level indicator stationary (when
Div
is selected) or to have it move with
the actual voltage level (when
Volts
is selected). The advantage of selecting
Div
is that the waveform remains on the grid as you increase the gain. If
Volts
is selected, the waveform could move off the grid.
NOTE:
Regardless of whether you select Volts or Div, the
Offset
shown in the
channel setup dialog always indicates volts. However, when Div is selected for
the Offset Control, the offset in volts is scaled proportional to the change in
gain, thereby keeping the division on the grid constant.
Touch
Utilities
Preferences
from the menu bar.
Touch the
Acquisition
tab.
Under
Offset Setting constant in
:, touch either the
Div
or
Volts
button.
Delay Control
As you change the timebase, this control allows you to either keep the
horizontal offset indicator stationary (when
Div
is selected) or to have it move
with the trigger point (when
Time
is selected). The advantage of selecting
Div
is that the trigger point remains on the grid as you increase the timebase. If
Time
is selected, the trigger point could move off the grid.
NOTE:
Regardless of whether you select Time or Div, the
Delay
shown in the
timebase setup dialog always indicates time. However, when Div is selected
for Delay In, the delay in time is scaled proportional to the change in timebase,
thereby keeping the division on the grid constant.
Touch
Utilities
Preferences
from the menu bar.
Summary of Contents for waverunner mxi-a series
Page 1: ...Getting Started Manual WaveRunner Xi A WaveRunner MXi A Series Oscilloscopes...
Page 2: ......
Page 3: ...WaveRunner Xi A MXi A Series Oscilloscopes Getting Started Manual January 2013...
Page 22: ...WaveRunner Xi A and MXi A Oscilloscopes 14 922138 00 Rev A Front Panel Controls...
Page 126: ...WaveRunner Xi A and MXi A Oscilloscopes 118 922138 00 Rev A...
Page 127: ......
Page 128: ......