WaveRunner Xi-A and MXi-A Oscilloscopes
36
922138-00 Rev A
Vertical
:
Turn the
L
EVEL
knob in the TRIGGER control group to
adjust the vertical threshold of the trigger or the
highlighted trace. Level defines the source voltage at
which the trigger generates an event: a change in the
input signal that satisfies the trigger conditions.
Alternatively, on the
Trigger
dialog, touch the
Level
field and provide a value (using the pop-up numeric
keypad).
Quickly set a level of zero volts by pushing the
Level
button on the front panel.
Edge Trigger Setup
Push the front panel trigger
S
ETUP
button.
Touch the
Edge
trigger button.
Touch inside the
Source
field and select an input
from the pop-up menu.
C1
through
C4
are abbreviations for Channel 1
through Channel 4.
Ext
and
Ext/10
allow you to trigger on the External
Input of the oscilloscope.
Ext
input is +/-0.5 V.
Ext/10
input is
+/-5.0 V.
Line
is for triggering on the Positive or Negative