Getting Started Manual
922138-00 Rev A
99
B
application. This is done using a recovery routine that restores a clean copy of
the image originally installed on the C: drive.
NOTE:
Any user and calibration data located on the D: partition is not affected
by the recovery process.
Teledyne LeCroy provides you with a recovery application and a backup image
in an extra partition on the instrument’s hard drive. The recovery process is
easy to perform, using the Recovery Wizard. After the recovery procedure is
done, you must activate Windows, either by Internet connection to
Microsoft’s Web site or by telephone. Have your Windows Product Key
number (located on the rear of the oscilloscope) handy during Widows
reactivation.
NOTE:
The recovery process produces a replica of the operating system and
oscilloscope application software at the current revision levels when the
oscilloscope was manufactured. Any further revisions of the application
software, Windows operating system, and virus scan definition files are not
automatically upgraded. Therefore, after completion of the disk image
recovery, it is highly recommended to search vendor Web sites and upgrade
the individual components to current revision levels.
The current oscilloscope application software can be downloaded directly
from the Teledyne LeCroy website at
teledynelecroy.com
. Since the calibration
data for the oscilloscope is stored in the D: drive, current calibration constants
are not overwritten during the recovery process.
Recovery Procedure
Connect a network cable to the LAN port on the rear of the oscilloscope if you
intend to activate windows through the Internet.
Connect a keyboard and a mouse to the oscilloscope.
Apply power to the oscilloscope.
As soon as the
Starting Acronis Loader... Press F11 for Acronis Startup
Recovery Manager
message appears on the screen, press the
F11
key until the
recovery software logo appears momentarily. The
Acronis True Image Echo
Workstation
main window is displayed.
NOTE:
Do not press
F11
before you see the
Starting Acronis Loader
message
or you will enter the MB boot sequence selection.
Select Recovery from the Main Menu. The Restore Data Wizard opens.
Summary of Contents for waverunner mxi-a series
Page 1: ...Getting Started Manual WaveRunner Xi A WaveRunner MXi A Series Oscilloscopes...
Page 2: ......
Page 3: ...WaveRunner Xi A MXi A Series Oscilloscopes Getting Started Manual January 2013...
Page 22: ...WaveRunner Xi A and MXi A Oscilloscopes 14 922138 00 Rev A Front Panel Controls...
Page 126: ...WaveRunner Xi A and MXi A Oscilloscopes 118 922138 00 Rev A...
Page 127: ......
Page 128: ......