Chapter 4 Common troubleshooting
Here are some common failures and their solutions. If the problem persists
after following the listed steps, please contact
Teledyne LeCroy
.
1. The instrument will not start up.
• Check whether the power is correctly connected.
• Check whether the power switch at the front panel is on.
• Remove the power cord and check whether the voltage selector is on
the correct scale, whether the specification of the fuse is correct and
whether the fuse is intact. If the fuse needs to be changed, please refer
to “
To Replace the Fuse
”.
• If the problem remains, please contact
Teledyne LeCroy
.
2. The constant voltage output is abnormal.
• Check whether the maximum output power of the scale currently
selected fulfills the load requirement. If yes, go to the next step.
• Check the cable connecting the load and power supply for short-
circuits.
• Check whether the load is normal.
• Check whether the current setting value of this scale is proper. If it is
too low, increase it.
• If the problem remains, please contact
Teledyne LeCroy
.
3. The constant current output is abnormal.
• Check whether the maximum output power of the scale currently
selected fulfills the load requirement. If yes, go to the next step.
• Check whether the cable connecting the load and power supply is in
good condition.
•
Check whether the load is normal.
•
Check whether the voltage setting value of this scale is proper. If it is
too low, increase it.
•
If the problem remains, please contact
Teledyne LeCroy
.