1.6 Output Inspection
1. Check the output voltage
•
Turn on the power and make sure the channel current setting is not zero
when the instrument has no-load.
•
Press on/off button, the supply should be working in constant voltage
(CV) mode. You can set the voltage range of the T3PS16081P by
adjusting the voltage setpoint from the minimum (0 V) to the maximum
value (16 V) and the voltage range of the T3PS30051P by adjusting the
voltage setpoint from the minimum (0 V) to the maximum value (30 V).
2. Check the output current
•
Turn on the power and make sure the voltage setting is not zero.
•
Connect the output terminals (short) with an insulated wire that can handle
10 A or more (18 AWG single core, for example).
•
Activate the output by pressing the on/off button. The low impedance
(shorted) output will cause the instrument to enter current control (CC)
mode. You can set the current range of the T3PS16081P by adjusting the
current setpoint from the minimum (0 A) to the maximum value (8 A) and the
current range of the T3PS30051P by adjusting the current setpoint from the
minimum (0 A) to the maximum value (5 A).