38
Table 2-6 Menu Explanations of Arb Waveform
Function menu
Explanations
SRate / Frequency
Set the signal sampling rate or frequency;
The current parameter will be switched at a second press
Amplitude / HighLevel
Set the signal amplitude or high level;
The current parameter will be switched at a second press
Offset / LowLevel
Set the signal offset or low level;
The current parameter will be switched at a second press
Arb Mode
Set the output mode to TrueArb or DDS
Arb Type
Select either built-in or stored waveforms
Page 1/2
Enter the next page
Phase
Set the phase of the signal
Interpolation
Set the interpolation mode to 0-order hold or Linear
Page 2/2
Return to the previous page
To Set the Sampling Rate
In TrueArb output mode, users can set the sampling rate (the output points-
per-second) or frequency of the arbitrary waveform. The instrument outputs an
arbitrary waveform point-by-point with the current sampling rate.
Press Waveforms → Page 1/2 → Arb → TrueArb → SRate, to set the sampling
rate parameter. When changing parameter, if the new value is valid, the value
will be set. Otherwise, the nearest valid value will be set.
To Set the Interpolation Mode
In TrueArb output mode, the T3AFG supports two interpolation modes of
including 0-order hold and Linear. Users can set the parameter by pressing
Interpolation in the operation menu. Figure 2-19 shows the output waveform
comparison of a 32-point sinusoidal arbitrary wave with "0-order hold"
interpolation and "linear" interpolation.