Foxy
®
R1 and Foxy
®
R2 Fraction Collectors
Section 4 Serial Command Control
4-14
THRESHOLD=n
This command sets the peak threshold (
n=0
[none] to 100)% of the current program.
The
THRESHOLD
command without an operand returns the
current setting.
WIDTH=n
This command sets the peak detection width of the
current program. Choices are:
•
n=1
5 seconds
•
n=2
15 seconds
•
n=3
30 seconds
•
n=4
1 minute
•
n=5
2 minutes
•
n=6
4 minutes
•
n=7
8 minutes
•
n=8
16 minutes
•
n=9
32 minutes
•
n=10
1 hour
•
n=11
No slope detection
The
WIDTH
command without an operand returns the current
setting.
WINDOW=n
The maximum number of windows is four. This command sets
which window number (
n=1
) will be used for the
WSTART
and
WEND
commands that follow.
Example:
Command:
THRESHOLD=10
Response:
(none)
Example:
Command:
THRESHOLD
Response:
THRESHOLD=10
Example:
Command:
WIDTH=2
Response:
(none)
Example:
Command:
WIDTH
Response:
WIDTH=2
Example:
Command:
WINDOW=1
Response:
(none)
Summary of Contents for Foxy R1
Page 2: ......
Page 4: ......
Page 12: ...Foxy R1 and Foxy R2 Fraction Collectors Safety xii ...
Page 24: ...Foxy R1 and Foxy R2 Fraction Collectors Section 1 Introduction 1 12 ...
Page 38: ...Foxy R1 and Foxy R2 Fraction Collectors Section 2 Preparation for Use 2 14 ...
Page 60: ...Foxy R1 and Foxy R2 Fraction Collectors Section 3 Programming 3 22 ...
Page 78: ...Foxy R1 and Foxy R2 Fraction Collectors Section 4 Serial Command Control 4 18 ...
Page 92: ...Foxy R1 and Foxy R2 Fraction Collectors Appendix A Icon Glossary and Menu Charts A 10 ...
Page 96: ...Foxy R1 and Foxy R2 Fraction Collectors Index Index 4 ...
Page 98: ......
Page 100: ......
Page 102: ......