Foxy
®
R1 and Foxy
®
R2 Fraction Collectors
Section 4 Serial Command Control
4-13
The
RSTIME
command without an operand returns the current
setting.
R1LAST=n or R2LAST=n
These commands set the last tube
for Rack1 or Rack2. Choices are
n=1
to
n=maximum
number of
tubes for programmed rack type.
The
R1LAST or R2LAST
command without an operand returns
the current setting.
R1TYPE=n or R2TYPE=n
These commands set the rack
type of the current program. Choices are:
•
n=0
none
•
n=1
12/13 mm tubes
•
n=2
16 mm tubes
•
n=3
18 mm tubes
•
n=4
17 mm vials
•
n=5
25/28 mm vials
•
n=6
1.5 mL microcentrifuge tubes
•
n=7
36 funnel prep. rack
•
n=8
Bottles
•
n=9
Single Beaker
•
n=10
Microplates, 96 well
2
•
n=11
Microplates, 96 well
6
•
n=12
26 funnel prep. rack
The
RTYPE
command without an operand returns the current
setting.
Example:
Command:
RSTIME
Response:
RSTIME=5000
Example:
Command:
R2LAST=10
Response:
(none)
Example:
Command:
R2LAST
Response:
R2LAST=10
Example:
Command:
RTYPE=5
Response:
(none)
Example:
Command:
RTYPE
Response:
RTYPE=5
Summary of Contents for Foxy R1
Page 2: ......
Page 4: ......
Page 12: ...Foxy R1 and Foxy R2 Fraction Collectors Safety xii ...
Page 24: ...Foxy R1 and Foxy R2 Fraction Collectors Section 1 Introduction 1 12 ...
Page 38: ...Foxy R1 and Foxy R2 Fraction Collectors Section 2 Preparation for Use 2 14 ...
Page 60: ...Foxy R1 and Foxy R2 Fraction Collectors Section 3 Programming 3 22 ...
Page 78: ...Foxy R1 and Foxy R2 Fraction Collectors Section 4 Serial Command Control 4 18 ...
Page 92: ...Foxy R1 and Foxy R2 Fraction Collectors Appendix A Icon Glossary and Menu Charts A 10 ...
Page 96: ...Foxy R1 and Foxy R2 Fraction Collectors Index Index 4 ...
Page 98: ......
Page 100: ......
Page 102: ......