Foxy
®
R1 and Foxy
®
R2 Fraction Collectors
Section 4 Serial Command Control
4-12
RANGE=n
This command sets the detector signal voltage range of the
current program. The choices are:
•
n=1
1 Volt
•
n=2
100 mV
•
n=3
10 mV
•
n=4
External detector
The
RANGE
command without an operand returns the current
setting.
RESTART=n
This command sets the restart action of the
current program.
Choices are:
•
n=1
None
•
n=2
Inject, Overlay
•
n=3
Inject, Skip Tube
•
n=4
Time, Overlay
•
n=5
Time, Skip Tube
•
n=6
Inject, Next Tube
•
n=7
Time, Next Tube
The
RESTART
command without an operand returns the current
setting.
RSTIME=n
This command sets the restart time (
n=0
to
359999
seconds) of the current program.
Example:
Command:
RANGE=2
Response:
(none)
Example:
Command:
RANGE
Response:
RANGE=2
Example:
Command:
RESTART=5
Response:
(none)
Example:
Command:
RESTART
Response:
RESTART=5
Example:
Command:
RSTIME=600
Response:
(none)
Summary of Contents for Foxy R1
Page 2: ......
Page 4: ......
Page 12: ...Foxy R1 and Foxy R2 Fraction Collectors Safety xii ...
Page 24: ...Foxy R1 and Foxy R2 Fraction Collectors Section 1 Introduction 1 12 ...
Page 38: ...Foxy R1 and Foxy R2 Fraction Collectors Section 2 Preparation for Use 2 14 ...
Page 60: ...Foxy R1 and Foxy R2 Fraction Collectors Section 3 Programming 3 22 ...
Page 78: ...Foxy R1 and Foxy R2 Fraction Collectors Section 4 Serial Command Control 4 18 ...
Page 92: ...Foxy R1 and Foxy R2 Fraction Collectors Appendix A Icon Glossary and Menu Charts A 10 ...
Page 96: ...Foxy R1 and Foxy R2 Fraction Collectors Index Index 4 ...
Page 98: ......
Page 100: ......
Page 102: ......