Operation
3000ZA
Teledyne Analytical Instruments
48
L—10 M—100
H—250 Mode—AUTO
Use the __ arrow keys to blink the range to be set: low (L), medium
(M), or high (H).
Use the __ arrow keys to enter the upper value of the range (all
ranges begin at 0 ppm). Repeat for each range you want to set. Press
Enter
to accept the values and return to Analyze mode. (See note
below.)
Note: The ranges must be increasing from low to high, for
example, if range 1 is set as 0–10 ppm and range 2 is set
as 0–100 ppm, range 3 cannot be set as 0– 50 ppm since
it is lower than range 2.
4.6.2
Fixed Range Analysis
The autoranging mode of the instrument can be overridden, forcing
the analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, enter the range
function by pressing the
Range
button on the front panel.
Use the __ arrow keys to move the blinking over
AUTO
.
Use the __ arrow keys to switch from
AUTO
to
FX/LO
,
FX/MED
, or
FX/HI
to set the instrument on the desired fixed range (low, medium, or
high).
L—10 M—100
H—250 Mode—FX/LO
—or—
L—10 M—10 0
H—250 Mode—FX/MED
—or—
L—10 M—10 0
H—250 Mode—FX/HI
Press
Escape
to re-enter the Analyze mode using the fixed range.
Note: When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit (or default value)
Summary of Contents for 3000ZA
Page 24: ...Operational Theory 3000ZA Teledyne Analytical Instruments 12 Figure 2 3 Component Location...
Page 63: ...Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 51...
Page 77: ...Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 1 SCFH 0 476 SLPM...
Page 81: ...Trace Oxygen Analyzer Index Teledyne Analytical Instruments...