Trace Oxygen Analyzer
Operation
Teledyne Analytical Instruments
33
The factory default is three minutes, but the delay time is
programmable. To adjust to delay time use the Left or Right arrow keys.
When the time displayed on the second line blinks, it can be adjusted by
Pressing the Up or Down keys to increase or decrease its value. The
minimum delay is 1 minute, the maximum is 30.
This preference is stored in non-volatile memory so that it is
recovered if power is removed from the instrument.
4.3.2
Setting up an Auto-Cal
When proper automatic valving is connected (see Chapter 3,
Installation
), the Analyzer can cycle itself through a sequence of steps
that automatically calibrates the instrument.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000ZA is acurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
To setup an Auto–Cal cycle:
Choose
System
from the Function buttons. The LCD will display
five sub functions.
TRAK/HLD Auto—Cal
PSWD Logout More
Use __ arrows to blink
Auto—Cal
, and press
Enter
. A new screen
for
Span/Zero
set appears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
Press __ arrows to blink Span (or Zero), then press
Enter
again.
(You won’t be able to set OFF to ON if a zero interval is entered.) A
Span Every ...
(or
Zero Every ...
) screen appears.
Span Every 0 d
Start 0 h from now
Use __ arrows to set an interval value, then use __ arrows to move
to the start-time value. Use__ arrows to set a start-time value.
To turn ON the Span and/or Zero cycles (to activate Auto-Cal):
Press
System
again, choose
Auto—Cal
, and press
Enter
again. When the
Span/Zero values screen appears, use the __ arrows to blink the Span (or
Summary of Contents for 3000ZA