Example
Calibration of an “acetone” detector using a calibration gas with 1% butane
Value to be displayed:
1% (injected butane) x 100 x 0.90 (butane/acetone coefficient) = 60% LEL
1.5% (butane LEL)
Note:
•
LELs vary according to the source.
•
Coefficients are accurate to ± 15%.
9.4
Maintaining the remote control
9.4.1
Replacing the batteries
The two AA batteries (1.5 V) in the device need to be replaced if transmission quality decreases.
In this case, remove the remote control (1) from its case (2), remove the cover from the battery
compartment (4) and replace the old batteries (3) with two new identical batteries. Replace the
cover (4), insert and tighten the screws, and put the remote control (1) back into its case (2).
Figure 64: installing new AA batteries (1.5 V).
Summary of Contents for OLCT 80 d
Page 1: ...NPO80GB Revision F 1 OLCT 80 GAS TRANSMITTER USER MANUAL...
Page 10: ...OLCT 80 GAS TRANSMITTER USER MANUAL 4 NPO80GB Revision F 1...
Page 56: ...OLCT 80 GAS TRANSMITTER USER MANUAL 50 NPO80GB Revision F 1...
Page 78: ...OLCT 80 GAS TRANSMITTER USER MANUAL 72 NPO80GB Revision F 1...
Page 80: ...OLCT 80 GAS TRANSMITTER USER MANUAL 74 NPO80GB Revision F 1...
Page 86: ...OLCT 80 GAS TRANSMITTER USER MANUAL 80 NPO80GB Revision F 1...
Page 93: ...OLCT 80 GAS TRANSMITTER USER MANUAL NPO80GB Revision F 1 87...
Page 94: ...OLCT 80 GAS TRANSMITTER USER MANUAL 88 NPO80GB Revision F 1...
Page 96: ...OLCT 80 GAS TRANSMITTER USER MANUAL 90 NPO80GB Revision F 1...
Page 104: ...OLCT 80 GAS TRANSMITTER USER MANUAL 98 NPO80GB Revision F 1...
Page 115: ...OLCT 80 GAS TRANSMITTER USER MANUAL NPO80GB Revision F 1 109...