6.11
Magnets operation
The digital display gives access to different menus.
To enter in these menus, the wand has to put in place on the stickers “PG1” or “PG2”. The user
must maintain the wand on the magnets with time longer or shorter depending on the desired
function. The following symbols that appear on the display give indication of time:
•
-: brief contact
•
<: short contact corresponding to 1s
•
<<: long contact corresponding to 2s
•
<<<: very long contact corresponding to 5s
For example:
To modify the calibration gas value, put the wand on the “PG2” sticker during 5 seconds until
the symbol <<< appears. The display indicates “VIEW CONFIG”. Put in place the wand on PG2
until “CALIB GAS” appears. Put the wand on PG1 during 1 second to enter in the menu.
Depending on gas concentration injected, decrease the value with PG1 or increase it with PG2.
In his normal state, the detector displays every 30 minutes the gas unit, the gas, the label and
the concentration. When the detector is in fault, the message “DEF” replaces the concentration.
When the detector is in alarm, la concentration stays on the display.
Some function requires a code that is active for 10 minutes. This maintenance code can be
change by the user (“0” means that the code is inactive, any value other than zero involves the
code request to access to the desired function).
These codes are saved in case of shutdown or reset.
The menu about the relays is available only If there is no optional card. Only the
relay state (energized / not energized) can be modified by the magnetic wand.
If you have a network card in the detector, the menu Alarm is unavailable.
Summary of Contents for Everywhereyoulook DG-TT7-K
Page 18: ...MULTIXPLO AND MULTITOX DETECTORS OPERATING MANUAL 12 NOSP17665 Revision 02a...
Page 34: ...MULTIXPLO AND MULTITOX DETECTORS OPERATING MANUAL 28 NOSP17665 Revision 02a...
Page 58: ...MULTIXPLO AND MULTITOX DETECTORS OPERATING MANUAL 52 NOSP17665 Revision 02a...
Page 66: ...MULTIXPLO AND MULTITOX DETECTORS OPERATING MANUAL 60 NOSP17665 Revision 02a...
Page 68: ...MULTIXPLO AND MULTITOX DETECTORS OPERATING MANUAL 62 NOSP17665 Revision 02a...
Page 69: ...MULTIXPLO AND MULTITOX DETECTORS OPERATING MANUAL NOSP17665 Revision 02a 63...
Page 70: ...MULTIXPLO AND MULTITOX DETECTORS OPERATING MANUAL 64 NOSP17665 Revision 02a...
Page 71: ...MULTIXPLO AND MULTITOX DETECTORS OPERATING MANUAL NOSP17665 Revision 02a 65...