Model T200U NOy Option Manual Addendum
34
07303E DCN7996
SECONDARY MENU (The fixed characteristics of the system).
PRESS
DISPLAY
ACTION
SEL
hold until
p-n1
SET TO 0 – REVERSE ACTION
*Thermocouple
SEL
hold until
Type K:
p-n2
SET TO 3 = TYPE K THERMOCOUPLE
*Thermocouple
SEL
hold until
Type J:
p-n2
SET TO 2 = TYPE J THERMOCOUPLE
SEL
p-dF
DIGITAL FILTER
SET TO 5
SEL
P-SL
LOWER LIMIT
SET TO -40 (-40ºC)
SEL
P-SU
UPPER LIMIT
SET TO 350 (350ºC)
SEL
P-AH
ALARM
SET TO 0
SEL
P-An
HYTERESIS
SET TO 5
SEL
P-dP
PROG DECIMAL
SET TO 0
SEL
PVOF
PROCESS OFFSET
LEAVE AT 0
SEL
SVOF
SET POINT OFFS
LEAVE AT 0
SEL
P-F
SET TO 0
(0 =
o
C, 1 =
o
F)
SEL
P-n1
ALREADY CONFIGURED
SEL
hold until
RETURNS TO MAIN MENU
SEL
PV/SV
SET DISPLAY TO 315
°
C
SEL
hold until
P
PROP BAND
ALREADY CONFIGURED
SEL
SCROLL TO
A7
SET TO 1 (AUTOTUNE )
* See Figure 7 to determine whether the thermocouple is Type K or Type J.
Summary of Contents for T200U NOy
Page 2: ......
Page 6: ...Model T200U NOy Option Manual Addendum iv 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...Model T200U NOy Option Manual Addendum 8 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...Model T200U NOy Option Manual Addendum 10 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 20: ...Model T200U NOy Option Manual Addendum 18 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 24: ...Model T200U NOy Option Manual Addendum 22 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 30: ...Model T200U NOy Option Manual Addendum 28 07303E DCN7996 Figure 5 Pneumatics...
Page 38: ...Model T200U NOy Option Manual Addendum 36 07303E DCN7996 Figure 8 Electrical Diagram...