TELEDYNE API T200U NOy Operator'S Manual Download Page 2

 

 

 

Summary of Contents for T200U NOy

Page 1: ...00U Addendum PN 06861 TELEDYNE API 9970 CARROLL CANYON ROAD SAN DIEGO CA 92131 1106 USA Toll free Phone 800 324 5190 Phone 1 858 657 9800 Fax 1 858 657 9816 Email api sales teledyne com Website http w...

Page 2: ......

Page 3: ...e instrument Technician Symbol All operations marked with this symbol are to be performed by qualified maintenance personnel only Electrical Ground This symbol inside the instrument marks the central...

Page 4: ...EN GARDE Surface chaude Ne pas toucher Toucher certaines parties de l instrument sans protection ou sans les outils appropri s pourrait entra ner des dommages aux pi ces ou l instrument Pictogramme t...

Page 5: ...provided but not manufactured by TAPI is warranted and will be repaired to the extent and according to the current terms and conditions of the respective equipment manufacturer s warranty PRODUCT RET...

Page 6: ...Model T200U NOy Option Manual Addendum iv 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...

Page 7: ...l Check Procedure 20 6 0 MAINTENANCE 23 6 1 Maintenance Schedule 23 6 2 Replacing the Sample Particulate Filters 24 6 3 Checking Analyzer Flow Rate 25 6 4 Replacing the Converter 26 6 5 Inspecting Pne...

Page 8: ...26 Figure 5 Pneumatics 28 Figure 6 Location of O3 Generator Air Inlet 29 Figure 7 Thermocouples J Type and K Type 35 Figure 8 Electrical Diagram 36 List of Tables Table 1 Specifications for the Exter...

Page 9: ...p The pump pack pulls ambient air into the converter box via the sample port the sample port s cross fitting splits the sample between two paths one bypasses the heated converter and flows directly in...

Page 10: ...Model T200U NOy Option Manual Addendum 8 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...

Page 11: ...feet 15 m maximum length Power 501 NOy Bypass Pump Assembly 120V 50 60 Hz 230V 50 60 Hz 360 watts Power Converter Assembly 120V 50 60 Hz 230V 50 60 Hz 60 watts Converter Efficiency 96 for NO2 Weight...

Page 12: ...Model T200U NOy Option Manual Addendum 10 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...

Page 13: ...ical cable assembly herein referred to as Converter or external Converter External Pump herein referred to as external Pump not to be confused with the 501 NOy Bypass Pump or Pump Pack 1 Verify that t...

Page 14: ...CODE NEC The umbilical cable assembly installation must follow local and national electrical code standards 3 Uncoil and arrange the TAPI provided umbilical cable assembly to avoid twisting or kinks w...

Page 15: ...prong thermocouple connector from the umbilical cable assembly to the THERMOCOUPLE connection on the rear of the 501 NOy 12 Connect the analyzer and 501 NOy AC power cords from their AC power recepta...

Page 16: ...Model T200U NOy Option Manual Addendum 14 07303E DCN7996 Figure 1 Rear Panels T200U NOy Analyzer top and 501 NOy center Converter bottom...

Page 17: ...nd rear panel labels include numbers to facilitate matching connections between instruments CONNECTIONS Analyzer to Pump to Converter and Calibrator to Pump Calibrator Analyzer Pump Chassis Converter...

Page 18: ...about 30 minutes for the external Converter to come up to operating temperature Caution The internal Converter components will become extremely hot and must be allowed to cool after power off before...

Page 19: ...xternal Converter and measures NO in the sample The other pulls the sample through the external Converter and is displayed as NOy Under normal calibration conditions the NOy measurement is equivalent...

Page 20: ...Model T200U NOy Option Manual Addendum 18 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...

Page 21: ...ration gas being provided to the 501 NOy bypass system also flows to the analyzer Flow requirements are 0 80 LPM for NO bypass 0 80 LPM for NOy bypass and 1 LPM for the analyzer Calibration gas flow m...

Page 22: ...instrument 4 Press ENTR Pressing ENTR actually changes the calculation equations 5 Press EXIT The analyzer returns to sampling Immediately after calibration data are not added to the DAS averages Tab...

Page 23: ...reading to stabilize at span value 3 Press SPAN If performing only Span Cal Check do not press ENTR press EXIT to avoid actually spanning the instrument 4 Press ENTR Pressing ENTR actually changes th...

Page 24: ...Model T200U NOy Option Manual Addendum 22 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...

Page 25: ...s Daily Particulate Filter Weekly as needed Ozone Flow Check every year and replace o ring and orifice in reaction cell Ozone Dryer Filter DFU Replace every year Reaction Cell Window Clean annually or...

Page 26: ...3 for an exploded view of the filter assembly 3 Visually inspect the filter through the glass window 4 If the filter appears dirty unscrew the hold down ring remove the Teflon o ring and then the filt...

Page 27: ...servicing Check the Pump and pneumatic system for leaks or rebuild the Pump CAUTION Never operate the analyzer without the ozone destruct component properly seated and connected within the pneumatic...

Page 28: ...t the gas fittings the power connector and the thermocouple to the converter Figure 4 5 Remove the interior assembly box from the chassis by loosening the 4 captive screws that secure the assembly to...

Page 29: ...ll affect both flow rate and response time It is important that the pneumatic system be periodically inspected and thoroughly cleaned if necessary Clean by disassembling and passing methanol through t...

Page 30: ...Model T200U NOy Option Manual Addendum 28 07303E DCN7996 Figure 5 Pneumatics...

Page 31: ...rt of the 501 NOy and the ozone generator air inlet located inside the analyzer Figure 6 Figure 6 Location of O3 Generator Air Inlet 3 Insert a shut off valve between the external Pump and the EXHAUST...

Page 32: ...ker to the instrument exhaust port CAUTION Pressure must be less than 15 PSI 3 Pressurize the system and check for leaks by watching overall pressure The pressure should not drop more than 1 Hg 5psi i...

Page 33: ...ent does not respond to span gas check troubleshooting section in the T200 manual The above should get you started in diagnosing and repairing the most common faults If these reasons have been elimina...

Page 34: ...arameters have been set at the factory and no further adjustment should be necessary Should you need to adjust the temperature follow these steps 1 Select SV with the PV SV button 2 Press the up arrow...

Page 35: ...int LED turns off Autotune is complete No further adjustment should be necessary Table 6 Temperature Controller Programming Guide PRESS DISPLAY ACTION SEL DATA P PROP BAND UP DOWN SET TO 5 0 ENTER SEL...

Page 36: ...DIGITAL FILTER SET TO 5 SEL P SL LOWER LIMIT SET TO 40 40 C SEL P SU UPPER LIMIT SET TO 350 350 C SEL P AH ALARM SET TO 0 SEL P An HYTERESIS SET TO 5 SEL P dP PROG DECIMAL SET TO 0 SEL PVOF PROCESS OF...

Page 37: ...Model T200U NOy Option Manual Addendum 07303E DCN7996 35 Figure 7 Thermocouples J Type and K Type...

Page 38: ...Model T200U NOy Option Manual Addendum 36 07303E DCN7996 Figure 8 Electrical Diagram...