6.2 Replacing the Sample Particulate Filters
The particulate filter should be inspected often for signs of plugging or
contamination. It is also common for dirt particles to cause instrument drift, and
affect accuracy.
To check and change the filter:
1.
Fold down the 501 NOy front panel.
2. Locate both filters on the left and right side of the 501 NOy front panel. See
Figure 3 for an exploded view of the filter assembly.
3. Visually inspect the filter through the glass window.
4. If the filter appears dirty, unscrew the hold-down ring, remove the Teflon o-ring
and then the filter.
5. Replace the filter, being careful that the element is fully seated in the bottom
of the holder. Replace the Teflon o-ring (refer to Figure 3 for orientation of o-
ring with cutouts), then screw on the hold-down ring and hand tighten. Do not
overtighten; this can cause the glass window to crack!
NOTE
Handle the filter assembly wetted surfaces (glass window, Teflon holder) as little as possible
and clean with cloth prior to assembly
Summary of Contents for T200U NOy
Page 2: ......
Page 6: ...Model T200U NOy Option Manual Addendum iv 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...Model T200U NOy Option Manual Addendum 8 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...Model T200U NOy Option Manual Addendum 10 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 20: ...Model T200U NOy Option Manual Addendum 18 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 24: ...Model T200U NOy Option Manual Addendum 22 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 30: ...Model T200U NOy Option Manual Addendum 28 07303E DCN7996 Figure 5 Pneumatics...
Page 38: ...Model T200U NOy Option Manual Addendum 36 07303E DCN7996 Figure 8 Electrical Diagram...