Model T200U NOy Option Manual Addendum
07303E DCN7996
33
To start the Autotune procedure:
1.
Press SEL until A7 (AT) is displayed
2.
Press Data until value = 1 (enable Autotune)
•
While Autotuning, the decimal point LED will blink.
•
NOTE: During the procedure, it is normal for the temp to overshoot up to
100°C.
•
When the decimal point LED turns off, Autotune is complete. No further
adjustment should be necessary.
Table 6 Temperature Controller Programming Guide
PRESS
DISPLAY
ACTION
SEL
DATA
P
PROP BAND
UP/DOWN SET TO 5.0 (ENTER)
SEL
DATA
i
INTEGRAL
SET TO 50
SEL
DATA
d
DERIVATIVE TIME
SET TO 15
SEL
DATA
AL
LOW ALARM SETP
SET TO 300
SEL
DATA
AH
HIGH ALARM SETP
SET TO 330
SEL
DATA
7C
CYCLE TIME
SET TO 2 (FOR FAST SYSTEM)
SEL
HYS
HYSTERESIS
SET TO 0.3
SEL
LOC
LOCK
0 = OPEN 1 = LOCKED
2 = SV ONLY OPEN
NOTE:
DO NOT SET THE TEMPERATURE (SV) HIGHER THAN 320
o
C.
Summary of Contents for T200U NOy
Page 2: ......
Page 6: ...Model T200U NOy Option Manual Addendum iv 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...Model T200U NOy Option Manual Addendum 8 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...Model T200U NOy Option Manual Addendum 10 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 20: ...Model T200U NOy Option Manual Addendum 18 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 24: ...Model T200U NOy Option Manual Addendum 22 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 30: ...Model T200U NOy Option Manual Addendum 28 07303E DCN7996 Figure 5 Pneumatics...
Page 38: ...Model T200U NOy Option Manual Addendum 36 07303E DCN7996 Figure 8 Electrical Diagram...