Model T200U NOy Option Manual Addendum
07303E DCN7996
25
Figure 3. Replacing the Particulate Filter
6.3 Checking Analyzer Flow Rate
The External Pump is capable of maintaining the analyzer reaction cell pressure at
less than 5.0 In-Hg-A. If a higher pressure is noted, the Pump may need servicing.
Check the Pump and pneumatic system for leaks or rebuild the Pump.
CAUTION!
Never operate the analyzer without the ozone destruct component properly
seated and connected within the pneumatic path. The ozone destruct, which is
heated, is integrated into a metal enclosure, at the exhaust of the reaction cell,
inside the instrument.
The sample flow as measured at ports NO IN (1) and NOy IN (2) at the rear of the
analyzer should be 1000
±
100 cc/min.
See Figure 1 for component locations.
Summary of Contents for T200U NOy
Page 2: ......
Page 6: ...Model T200U NOy Option Manual Addendum iv 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...Model T200U NOy Option Manual Addendum 8 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...Model T200U NOy Option Manual Addendum 10 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 20: ...Model T200U NOy Option Manual Addendum 18 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 24: ...Model T200U NOy Option Manual Addendum 22 07303E DCN7996 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 30: ...Model T200U NOy Option Manual Addendum 28 07303E DCN7996 Figure 5 Pneumatics...
Page 38: ...Model T200U NOy Option Manual Addendum 36 07303E DCN7996 Figure 8 Electrical Diagram...