![TELEDYNE API 465H User Manual Download Page 11](http://html1.mh-extra.com/html/teledyne-api/465h/465h_user-manual_1080885011.webp)
06161H DCN7969
Teledyne API Model 465H O
3
Monitor User Manual
9
2.
SPECIFICATIONS AND AGENCY
APPROVALS
2.1.
Specifications
PARAMETER
DESCRIPTION
Measuring principle
UV absorption (Beer Lambert Law)
Ranges
0-5 WT% to 0-25 WT%
0-100 g/Nm
3
to 0-400 g/Nm
3
Measurement Units
WT%, g/Nm
3
Accuracy
± 1% of Full Scale
Precision/Repeatability
± 0.5% of Full Scale
Display Resolution
0.01 WT%, 0.1 g/Nm
3
Response Time (95%)
< 5 seconds to 95%
Compensation
Pressure, Temperature (NTP = 273.15K, 760 mmHg)
Gas Inlet Pressure Range
3.0-30.0 psig
Sample Flow Rate
0.2-2.0 LPM
Humidity Range
10-90% RH, non-condensing
Temperature Range
5-45
o
C
Dimensions
(H x W x D)
5.22” x 19.0" x 15.3" (3U RETMA Panel)
(133 mm x 483 mm x 388 mm)
In NEMA 4X enclosure, typical product envelope:
16.85" x 15.60 (±0.25)" x 6.65"
(428mm x 396 (±6)mm x 169mm)
Weight
Rack Mount – 13.6 lbs (6.17 kg)
in NEMA 4X enclosure – 15.3 lbs (6.94 kg)
Power
100-240 V~, 47-63 Hz, 74 W
Environmental
Installation Category II
Pollution Degree 2
For Indoor Use Only
Maximum Operating Altitude 2000 meters
Analog Output (Bi-polar)
0-5V, 4-20 mA isolated output
Status Outputs
System OK, Invalid Reading, Check Lamp, Pneumatic Error
RS-232 (optional RS-485)
57.6 Kbaud, DB-9 connector
Summary of Contents for 465H