Incoming Inspection
Simultaneous Input
Operation Test (Option
SIM)
1.
Set up a signal generator with an HDVG1 module and a DVG1 module.
Connect an output from the HDVG1 module to the SDI A input on the
instrument under test.
2.
Connect an output from the DVG1 module to the SDI B input on the
instrument under test.
3.
Connect the loopthrough terminators on the instrument.
4.
Set the HDVG1 module format to 1080 59.94i 100% color bars.
5.
Set the DVG1 module format to 525-270.
6.
Press
FACTORY
to restore factory settings.
7.
Press
CONFIG
, select
Input Mode
, and then select
Simultaneous
.
8.
Press
CONFIG
again to remove the menu.
9.
Press
DISPLAY SELECT
button
2
to select tile 2.
10.
Press the
WFM
button to select the waveform display in tile 2.
11.
Press
DISPLAY SELECT
button
4
to select tile 4.
12.
Press the
PICT
button to select the picture display.
13.
Check that the left two tiles (tiles 1 and 3) show the signal from the SDI
A input. Check that the right two tiles (tiles 2 and 4) show the signal from
the SDI B input.
14.
Check that the left portion of the Status Bar displays the status of the SDI A
signal input, and that the right portion of the Status Bar displays the status of
the SDI B signal input.
15.
Enter Pass or fail in the test record.
Waveform Rasterizers Speci
fi
cations and Performance Veri
fi
cation
69