Incoming Inspection
Analog Audio Input (Option
AD & DDE)
1.
Press the FACTORY button to restore the factory preset.
2.
Set the analog audio generator to output a 1 kHz, 18dBu sine wave.
3.
Press and hold the
DISPLAY SELECT 4
button to display the AUDIO
pop-up menu.
4.
Select
Audio Input > Analog A
.
5.
Press the
AUDIO
button brie
fl
y to dismiss the menu.
6.
Using the audio breakout cable or equivalent, connect the
fi
rst Line A input
pair to the analog audio generator and verify that the output level of the
generator is indicated on the audio bars with an 18 dBu signal.
7.
Repeat step 6 for the second and third line A input pair.
8.
Record Pass or Fail in the test record.
9.
Press and hold the
Audio
button to display the AUDIO menu and select
Audio Input > Analog B
.
10.
Press the AUDIO button again to dismiss the menu.
11.
Repeat step 6 for all three Line B input pairs.
12.
Record Pass or Fail in the test record.