Incoming Inspection
18.
Set the SDI generator for embedded audio in 2 groups starting with group 1,
and then access and set its audio channels:
a.
Press Modules, and then select DVG1.
b.
Press Test Signals, and select Module Parameters
c.
Select Audio, and set the audio channels as follows:
Channel
Value
Channel 1
50 Hz, –35 dB
Channel 2
100 Hz, –30 dB
Channel 3
150 Hz, –25 dB
Channel 4
200 Hz, –20 dB
Channel 5
250 Hz, –15 dB
Channel 6
300 Hz, –10 dB
Channel 7
400 Hz, –5 dB
Channel 8
500 Hz, 0 dB
19.
Connect the output of the SDI signal generator with embedded audio to the
A SDI input.
20.
Press and hold the
DISPLAY SELECT 4
button to display the AUDIO
pop-up menu.
21.
Select
Audio Input > Embedded
.
22.
Press the
AUDIO
button to dismiss the menu.
23.
Verify that the level meter bars have a stair step pattern from –35 dB on
channel 1 to 0 dB on channel 8.
NOTE.
If level-meter bars indicate that audio is not present, set the
TG2000 module number of groups to 2 even if it appears to be set already.
24.
Record Pass or Fail for Embedded Display in the test record.
25.
Press and hold the
AUDIO
button.
26.
Select
Phase Pair 1 & 2
.
27.
Press the
Audio
button to close the menu.
28.
Verify that the phase display changes as appropriate.
29.
Repeat steps 25 through 27 for phase pairs 3 & 4, 5 & 6, and 7 & 8, and verify
that the phase display changes in each one.
30.
Record Pass or Fail for Embedded to Lissajous in the test record.
64
Waveform Rasterizers Speci
fi
cations and Performance Veri
fi
cation
Summary of Contents for WVR6020
Page 2: ......
Page 6: ......
Page 14: ...Preface viii Waveform Rasterizers Specifications and Performance Verification...