![Tektronix WVR6020 Technical Reference Download Page 106](http://html1.mh-extra.com/html/tektronix/wvr6020/wvr6020_technical-reference_1077619106.webp)
Video and General Performance Veri
fi
cation
Equalized Eye Gain
(Options EYE or PHY)
This test checks for the presence of an internally generated eye signal.
1.
Press FACTORY.
2. HD only:
Connect an HDVG1 output to the SDI A input and terminate the
loopthrough connectors. Set the HDVG1 to generate a 1080 59.94i 100%
Color Bars signal.
3. SD only
: Connect a DVG1 output to the SDI A input and terminate the
loopthrough connectors. Set the DVG1 to generate a 525-270 75% Color
Bars signal.
4.
Press the EYE button (or MEAS if no EYE button) to get the Eye display.
5.
Press CONFIG, and then select Physical Layer Settings > Eye Display Mode
> Equalized.
6.
Using the graticule, check that the waveform amplitude is between 400 mV
and 650 mV.
7.
Record Pass or fail in the test record.