Application Examples
You can see the
Δ
(delta) time and frequency (the measured ring frequency)
in the Cursor Menu.
Type
Time
Source
CH1
Δ
t 540.0ns
1/
Δ
t 1.852MHz
Δ
V 0.44V
Cursor1
180ns
1.40V
Cursor2
720ns
0.96V
8.
Push
Type
►
Amplitude
.
9.
Push the
Cursor 1
option .
10.
Turn the multipurpose knob to place a cursor on the
fi
rst peak of the ring.
11.
Push the
Cursor 2
option .
12.
Turn the multipurpose knob to place Cursor 2 on the lowest part of the ring.
You can see the amplitude of the ring in the Cursor Menu.
Type
Amplitude
Source
CH1
Δ
V 640mV
Cursor 1
1.46V
Cursor 2
820mV
Measuring Pulse Width
If you are analyzing a pulse waveform and you want to know the width of the
pulse, follow these steps:
1.
Push the
Cursor
to see the Cursor Menu.
2.
Push
Type
►
Time
.
3.
Push
Source
►
CH1
.
4.
Push the
Cursor 1
option .
TPS2000B Series Digital Oscilloscope User Manual
47
Summary of Contents for TPS2012B
Page 1: ...xx TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual P071273302 071 2733 02...
Page 2: ......
Page 3: ...TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual xx www tektronix com 071 2733 02...
Page 8: ......
Page 22: ...Compliance information xiv TPS2000B Series Digital Oscilloscope User Manual...
Page 26: ...Preface xviii TPS2000B Series Digital Oscilloscope User Manual...
Page 40: ...Getting Started 14 TPS2000B Series Digital Oscilloscope User Manual...
Page 93: ...Math FFT TPS2000B Series Digital Oscilloscope User Manual 67...
Page 96: ...Math FFT 70 TPS2000B Series Digital Oscilloscope User Manual...
Page 170: ...Appendix D Cleaning 144 TPS2000B Series Digital Oscilloscope User Manual...