Managing TPSBAT Battery Packs
Checking the Charge and Calibration Status
The
Utility
►
System Status
►
Misc
option shows the amount of time you can
continue to operate the oscilloscope from the battery packs and the battery charge
status. If the oscilloscope is operating from the AC adapter, only the battery
charge status is reported.
The remaining time calculation is based on a one minute running average of the
current drain for the battery pack in use. Wait at least one minute after you power
on the oscilloscope to ensure an accurate reading.
Calibration refers to the ability of a battery pack to report its current capacity,
which the oscilloscope uses to calculate how much longer it can operate under
the current conditions.
A battery pack loses its calibration after repeated cycles of discharging and
charging in which the battery pack is not allowed to fully discharge. For example,
if you operate a battery pack for one hour a day or less, and recharge it after each
use, over time the battery pack will become uncalibrated.
An uncalibrated battery pack cannot accurately report its remaining operating
time. (See page 90,
TPS2000B Series Digital Oscilloscope User Manual
87
Summary of Contents for TPS2012B
Page 1: ...xx TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual P071273302 071 2733 02...
Page 2: ......
Page 3: ...TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual xx www tektronix com 071 2733 02...
Page 8: ......
Page 22: ...Compliance information xiv TPS2000B Series Digital Oscilloscope User Manual...
Page 26: ...Preface xviii TPS2000B Series Digital Oscilloscope User Manual...
Page 40: ...Getting Started 14 TPS2000B Series Digital Oscilloscope User Manual...
Page 93: ...Math FFT TPS2000B Series Digital Oscilloscope User Manual 67...
Page 96: ...Math FFT 70 TPS2000B Series Digital Oscilloscope User Manual...
Page 170: ...Appendix D Cleaning 144 TPS2000B Series Digital Oscilloscope User Manual...