Appendix A: TPS2000B Speci
fi
cations
Table 5: Trigger Speci
fi
cations (cont.)
Characteristics
Description
Signal Formats and Field
Rates, Video Trigger Type
Supports NTSC, PAL, and SECAM broadcast systems for any
fi
eld or any line
Holdoff Range
500 ns to 10 s
1
Available only on a 4-channel oscilloscope.
2
TPS2024B only.
Table 6: Pulse Width Trigger Speci
fi
cations
Characteristics
Description
Pulse Width Trigger modes
Trigger when < (Less than), > (Greater than), = (Equal), or
≠
(Not Equal); Positive pulse or
Negative pulse
Pulse Width Trigger Point
Equal: The oscilloscope triggers when the trailing edge of the pulse crosses the trigger level.
Not Equal: If the pulse is narrower than the speci
fi
ed width, the trigger point is the trailing edge.
Otherwise, the oscilloscope triggers when a pulse continues longer than the time speci
fi
ed as
the Pulse Width.
Less than: The trigger point is the trailing edge.
Greater than (also called time-out trigger): The oscilloscope triggers when a pulse continues
longer than the time speci
fi
ed as the Pulse Width.
Pulse Width Range
Selectable from 33 ns to 10 s
Pulse Width Resolution
16.5 ns or 1 part per thousand, whichever is larger
Equal Guardband
t > 330 ns: ±5%
≤
guardband < ±(5.1% + 16.5 ns)
t
≤
330 ns: guardband = ±16.5 ns
Not Equal Guardband
t > 330 ns: ±5%
≤
guardband < ±(5.1% + 16.5 ns)
165 ns < t
≤
330 ns: guardband = -16.5 ns ÷ +33 ns
t
≤
165 ns: guardband = ±16.5 ns
Table 7: Trigger Frequency Counter Speci
fi
cations
Characteristics
Description
Readout Resolution
6 digits
Accuracy (typical)
±51 parts per million including all frequency reference errors and ±1 count errors
Frequency Range
AC coupled, 10 Hz minimum to rated bandwidth
Signal Source
Pulse Width or Edge Trigger modes: all available trigger sources
The Frequency Counter measures trigger source at all times in Pulse Width or Edge modes,
including when the oscilloscope acquisition is halted due to changes in the run status, or acquisition
of a single shot event has completed.
Pulse Width Trigger mode: The oscilloscope counts pulses of signi
fi
cant magnitude inside the
250 ms measurement window that qualify as triggerable events, such as narrow pulses in a PWM
pulse train if set to < mode and the width is set to a relatively small time.
Edge Trigger mode: The oscilloscope counts all edges of suf
fi
cient magnitude and correct polarity.
Video Trigger mode: The Frequency Counter does not operate.
TPS2000B Series Digital Oscilloscope User Manual
127
Summary of Contents for TPS2012B
Page 1: ...xx TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual P071273302 071 2733 02...
Page 2: ......
Page 3: ...TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual xx www tektronix com 071 2733 02...
Page 8: ......
Page 22: ...Compliance information xiv TPS2000B Series Digital Oscilloscope User Manual...
Page 26: ...Preface xviii TPS2000B Series Digital Oscilloscope User Manual...
Page 40: ...Getting Started 14 TPS2000B Series Digital Oscilloscope User Manual...
Page 93: ...Math FFT TPS2000B Series Digital Oscilloscope User Manual 67...
Page 96: ...Math FFT 70 TPS2000B Series Digital Oscilloscope User Manual...
Page 170: ...Appendix D Cleaning 144 TPS2000B Series Digital Oscilloscope User Manual...