Getting Started
Current Probe Scaling
Current probes provide a voltage signal proportional to the current. You need to
set the oscilloscope to match the scale of your current probe. The default scale is
10 A/V.
To set the scale, follow these steps:
1.
Push a vertical channel button (such as the Channel 1
Menu
button).
2.
Push the
Probe
option button.
3.
Push the
Current
option button.
4.
Push the
Scale
option button to select an appropriate value.
Self Calibration
The self calibration routine lets you optimize the oscilloscope signal path for
maximum measurement accuracy. You can run the routine at any time but you
should always run the routine if the ambient temperature changes by 5 °C (9 °F)
or more. The routine takes about two minutes.
For accurate calibration, power on the oscilloscope and wait twenty minutes to
ensure it is warmed up.
To compensate the signal path, disconnect any probes or cables from the input
connectors. Then, access the
Utility
►
Do Self Cal
option, and follow the
directions on the screen.
TPS2000B Series Digital Oscilloscope User Manual
13
Summary of Contents for TPS2012B
Page 1: ...xx TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual P071273302 071 2733 02...
Page 2: ......
Page 3: ...TPS2000B Series Digital Storage Oscilloscope ZZZ User Manual xx www tektronix com 071 2733 02...
Page 8: ......
Page 22: ...Compliance information xiv TPS2000B Series Digital Oscilloscope User Manual...
Page 26: ...Preface xviii TPS2000B Series Digital Oscilloscope User Manual...
Page 40: ...Getting Started 14 TPS2000B Series Digital Oscilloscope User Manual...
Page 93: ...Math FFT TPS2000B Series Digital Oscilloscope User Manual 67...
Page 96: ...Math FFT 70 TPS2000B Series Digital Oscilloscope User Manual...
Page 170: ...Appendix D Cleaning 144 TPS2000B Series Digital Oscilloscope User Manual...