Maintenance
6--26
TPS2000 Series Digital Storage Oscilloscopes Service Manual
Table 6--4: List of error codes (cont.)
Number
Probable cause
Description
ID
12
Diagnostics
CH3 diagnostics failed
Acquisition board or power
supply failure
13
Diagnostics
CH4 diagnostics failed
Acquisition board or power
supply failure
14
Diagnostics
ADG421 #2 diagnostics
failed
Acquisition board or power
supply failure
255
Software
Software defect
Contact Tektronix
Clearing the Error Log.
To clear the error log after fixing a problem, follow these
steps:
1.
Enable the Service menus as described on page 5--2.
2.
Push the
UTILITY
button and push the
Service
option button to display the
Service menu.
3.
Push the
Service Diag.
option button to display the Diagnostics menu.
4.
Push the
Clear Error Log
option button to clear the error log of past
failures. This also sets the Power Cycle count to zero.
Summary of Contents for TPS2000 Series
Page 6: ......
Page 12: ...Table of Contents vi TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 16: ...Service Safety Summary x TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 20: ...Preface xiv TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 42: ...Theory of Operation 3 6 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 62: ...Adjustment Procedures 5 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 90: ...Maintenance 6 28 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 92: ...Options 7 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 94: ...Electrical Parts List 8 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 96: ...Diagrams 9 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 106: ...Replaceable Parts 10 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...