Operating Information
2--4
TPS2000 Series Digital Storage Oscilloscopes Service Manual
Self Calibration
The self-calibration routine lets you quickly optimize the oscilloscope signal path
for maximum measurement accuracy. You can run the routine at any time but you
should always run the routine if the ambient temperature changes by 5
_
C or
more.
To compensate the signal path, disconnect any probes or cables from the channel
input connectors. Then, press the
UTILITY
button and select the
Do Self Cal
option. Follow the on-screen instructions to proceed.
Default Setup
Check the TPS2000 User Manual for the state of the oscilloscope after you press
DEFAULT SETUP
.