Adjustment Procedures
5- 2
TPS2000 Series Digital Storage Oscilloscopes Service Manual
Adjustment Procedure
The adjustment procedure consists of multiple steps for the two-channel
oscilloscope, and additional steps for the four-channel model. Each step requires
an external voltage source for new calibration constants.
Before performing the adjustment procedure, you must warm up the oscilloscope
for at least ten minutes in an ambient temperature between 20
°
C and 30
°
C.
Adjustments performed prior to warm-up or outside this temperature range may
result in poor performance.
If all steps in the procedures are completed successfully, a “Pass” message is
displayed and the new calibration constants take affect. If any step fails, the
procedure is aborted and the current calibration is not affected. You can choose to
abort the procedures at any step by selecting FCAL ABORT without affecting
the current calibration.
NOTE
. While performing the adjustment procedure, keep in mind that some of the
steps take a significant amount of time (up to 5 minutes) to complete.
The equipment setups, shown in Figure 5--1, are required to complete the
adjustment procedure. Table 5--2 on page 5--5 lists the steps in the procedure
and the signal requirement for each step.
You must enable the Service menu to perform the adjustment procedure.
1.
Power on the oscilloscope.
2.
Push the front-panel
MEASURE
button to access the MEASURE menu.
3.
Push the top option button to access the Measure 1 menu.
4.
Push and hold the front-panel
SINGLE SEQ
button.
5.
Push and hold the front-panel
AUTOSET
button.
6.
Wait at least two seconds.
7.
Release the
SINGLE SEQ
button.
8.
Release the
AUTOSET
button. A message appears in the lower left corner
of the screen stating “Service mode ON.”
9.
Push the front-panel
UTILITY
button. The last item in the Utility menu is
now Service.
You may now begin executing the sequence of calibration steps required by your
oscilloscope and shown below. After the adjustment is complete, disable the
Service menu by pushing the UTILITY front-panel button, and the Service and
Service Mode Off option buttons.
Enable the Service Menu
Summary of Contents for TPS2000 Series
Page 6: ......
Page 12: ...Table of Contents vi TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 16: ...Service Safety Summary x TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 20: ...Preface xiv TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 42: ...Theory of Operation 3 6 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 62: ...Adjustment Procedures 5 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 90: ...Maintenance 6 28 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 92: ...Options 7 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 94: ...Electrical Parts List 8 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 96: ...Diagrams 9 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 106: ...Replaceable Parts 10 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...