Adjustment Procedures
5- 4
TPS2000 Series Digital Storage Oscilloscopes Service Manual
Use this procedure to load new calibration constants.
1.
Enable the Service menu as described on page 5--2.
2.
Push the
Service
option button.
3.
Push the
Factory Cal
option button.
4.
Push the
Warmup Timer
option button.
5.
Push the
Start Timer
option button to ensure a 10-minute warm up. After
10 minutes passes, the oscilloscope displays a large black or green cross.
6.
Press the front-panel
UTILITY
button and push the
Service
option button.
7.
Push the
Factory Cal
option button.
8.
Push the
FCAL INIT
option button to initiate the factory calibration
routine. You are now in step 1 of the procedure.
9.
As indicated by Step 1 in Table 5--2, connect a 0.0 V
DC
source to the channel
inputs.
10.
Press
FCAL STEP
to load the calibration constant for step 1.
11.
Follow the screen prompts along with Table 5--2 for the remainder of the
routine, applying the proper signal for each step.
NOTE
. During some steps, the oscilloscope may appear to be idle for several
minutes while it is processing information internally. You must wait for the
screen prompt to continue.
NOTE
. In Table 5--2, “BWL” stands for the bandwidth limit of the particular
oscilloscope. If the oscilloscope is a TPS2012 or a TPS2014, the oscilloscope
has a bandwidth limit of 100 MHz. If the oscilloscope is a TPS2024, the
bandwidth limit is 200 MHz.
If any step in the procedure fails, the procedure terminates. Failure will result
if an incorrect signal source is connected. If this happens, you must start the
procedure over from the beginning.
If the procedure completes successfully, a “Pass” message is displayed and
the new adjustment takes effect.
12.
Compensate the signal path by performing the self-calibration routine.
Disconnect any probes or cables from the channel input connectors. Then,
press the
UTILITY
button and select
Do Self Cal
to confirm that you are
ready to proceed.
Adjustment Procedure
Summary of Contents for TPS2000 Series
Page 6: ......
Page 12: ...Table of Contents vi TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 16: ...Service Safety Summary x TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 20: ...Preface xiv TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 42: ...Theory of Operation 3 6 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 62: ...Adjustment Procedures 5 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 90: ...Maintenance 6 28 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 92: ...Options 7 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 94: ...Electrical Parts List 8 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 96: ...Diagrams 9 2 TPS2000 Series Digital Storage Oscilloscopes Service Manual...
Page 106: ...Replaceable Parts 10 10 TPS2000 Series Digital Storage Oscilloscopes Service Manual...