Adjustment Procedures
5- 2
TLA7AXX/TLA7NAX Logic Analyzer Module Service Manual
To perform the self calibration procedure, do the following:
1.
Ensure that the instrument has had a 30-minute warm up before attempting
the self calibration, and that the logic analyzer application is running.
2.
Disconnect any probes connected to the logic analyzer module.
3.
Select Calibration and Diagnostics from the System menu.
4.
Select the Self Calibration tab page.
5.
Select the logic analyzer module.
6.
Click the Run button to start the self calibration.
The self calibration takes several minutes to complete, depending on the
number of channels in the module. Upon successfully completing the self
calibration, the module status changes from Running to Calibrated, and the
Date and Time field is set to the present.
Summary of Contents for TLA7AA Series
Page 5: ......
Page 11: ...Table of Contents vi TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 15: ...Service Safety Summary x TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 21: ...Preface xvi TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 25: ...Introduction xx TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 27: ...Specifications 1 2 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 43: ...Operating Information 2 16 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 49: ...Theory of Operation 3 6 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 53: ...Performance Verification 4 4 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 95: ...Troubleshooting 6 40 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 101: ...Options 7 4 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 103: ...Electrical Parts List 8 2 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...
Page 107: ...Diagrams 9 4 TLA7AXX TLA7NAX Logic Analyzer Module Service Manual...