![Tektronix TLA715 Series Service Manual Download Page 36](http://html1.mh-extra.com/html/tektronix/tla715-series/tla715-series_service-manual_1077833036.webp)
Specifications
TLA715 Dual Monitor Portable Mainframe Service Manual
1- 13
Table 1- 10: Certifications and compliances
Category
Standards or description
EC Declaration of Conformity --
EMC
Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EN 61326
EMC requirements for Class A electrical equipment for
measurement, control and laboratory use.
1
IEC 61000--4--2
Electrostatic discharge immunity (Performance criterion B)
IEC 61000--4--3
RF electromagnetic field immunity (Performance criterion A)
IEC 61000--4--4
Electrical fast transient / burst immunity (Performance criterion B)
IEC 61000--4--5
Power line surge immunity (Performance criterion B)
IEC 61000--4--6
Conducted RF immunity (Performance criterion A)
IEC 61000--4--11
Voltage dips and interruptions immunity (Performance criterion B)
EN 61000--3--2
AC power line harmonic emissions
Australia / New Zealand
Declaration of Conformity
Complies with EMC provision of Radiocommunications Act per the following standard(s):
Declaration of Conformity --
EMC
AS/NZS 2064.1/2
Industrial, Scientific, and Medical Equipment: 1992
EC Declaration of Conformity --
Low Voltage
Compliance was demonstrated to the following specification as listed in the Official Journal of the
European Communities:
g
Low Voltage Directive 73/23/EEC, amended by 93/68/EEC
EN 61010-1/A2:1995
Safety requirements for electrical equipment for measurement
control and laboratory use.
Canadian Certification
CAN/CSA C22.2 No. 1010.1
Safety requirements for electrical equipment for measurement,
control, and laboratory use.
Installation (Overvoltage)
Category
Terminals on this product may have different installation (overvoltage) category designations. The
installation categories are:
g y
CAT III
Distribution-level mains (usually permanently connected). Equipment at this level is
typically in a fixed industrial location.
CAT II
Local-level mains (wall sockets). Equipment at this level includes appliances, portable
tools, and similar products. Equipment is usually cord-connected.
CAT I
Secondary (signal level) or battery operated circuits of electronic equipment.
Pollution Degree
A measure of the contaminates that could occur in the environment around and within a product.
Typically the internal environment inside a product is considered to be the same as the external.
Products should be used only in the environment for which they are rated.
Pollution Degree 2
Normally only dry, nonconductive pollution occurs. Occasionally a
temporary conductivity that is caused by condensation must be
expected. This location is a typical office/home environment.
Temporary condensation occurs only when the product is out of
service.
1
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test
object.
Summary of Contents for TLA715 Series
Page 5: ......
Page 11: ...Table of Contents vi TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 15: ...Service Safety Summary x TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 19: ...Preface xiv TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 23: ...Introduction xviii TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 39: ...Specifications 1 16 TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 59: ...Performance Verification Procedures 4 8 TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 61: ......
Page 63: ...Adjustment Procedures 5 2 TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 67: ...Maintenance 6 4 TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 103: ...Removal and Installation Procedures 6 40 TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 119: ...Troubleshooting 6 56 TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 121: ...Repackaging Instructions 6 58 TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 125: ...Electrical Parts List 8 2 TLA715 Dual Monitor Portable Mainframe Service Manual...
Page 132: ...Mechanical Parts List TLA715 Dual Monitor Portable Mainframe Service Manual 10 5...
Page 146: ......