Performance veri
fi
cation
2.
Set the oscilloscope settings as follows:
Control
Setting
Vertical
200 mV/div
Horizontal
100 ns/div
Acquire
Sample
Trigger position
50%
Trigger slope
Rising Edge
Input impedance
1 M
Ω
Measure
Amplitude
3.
Load the factory default preset. (See page 20,
4.
Use the oscilloscope to measure that the signal amplitude is within the range
of 900 mV to 1100 mV for the AES 1+2 channel.
5.
Record the measured value for AES 1+2 in the test record.
6.
Use the oscilloscope to measure that the signal amplitude is within the range
of 900 mV to 1100 mV for the remaining AES channels as follows:
Change the BNC cable connection from the AES 1+2 connector to the
AES 3+4 connector and record the measured value for AES 3+4 in the
test record.
Change the BNC cable connection from the AES 3+4 connector to the
AES 5+6 connector and record the measured value for AES 5+6 in the
test record.
Change the BNC cable connection from the AES 5+6 connector to the
AES 7+8 connector and record the measured value for AES 7+8 in the
test record.
Change the BNC cable connection from the AES 7+8 connector to the
DARS connector and record the measured value for DARS in the test
record.
7.
Change the BNC cable connection from the DARS connector to the AES
1+2 connector.
62
SPG8000 Speci
fi
cations and Performance Veri
fi
cation