Performance veri
fi
cation
Sine amplitude test
Perform this procedure to check the sine amplitude.
1.
Connect the
BLACK 3
output on the SPG8000 to the oscilloscope, and
terminate the input with a 75
Ω
feed-through terminator.
Figure 10: Setup for sine amplitude test
2.
Press the
BLACK
button until
BLACK 3 : FORMAT
appears.
3.
Use the left (
◄
) arrow button to select
CW 10 MHz
.
4.
Press the
ENTER
button.
5.
Measure the amplitude of the sine wave on the oscilloscope.
6.
Record the result in the test record.
50
SPG8000 Speci
fi
cations and Performance Veri
fi
cation