Command Descriptions
2−72
TDS200/1000/2000 Series Oscilloscope Programmer Manual
DATa:STARt 10
specifies that the waveform transfer will begin with data point 10.
DATa:STARt?
might return
214
as the first waveform data point that will be
transferred.
CURVe?
DATa:STOP
Sets or queries the last data point in the waveform that will be
transferred when executing the CURVe? command. This lets you
transfer partial waveforms from the oscilloscope.
When using the CURVe command, the oscilloscope stops reading
data when there is no more data to read or when the 2500 data point
limit is reached.
Waveform
DATa:STOP <NR1>
DATa:STOP?
<NR1>
is an integer value that ranges from 1 to 2500, and specifies
the last data point that will be transferred. When DATa:STOP is less
than DATa:STARt, the values are swapped internally for CURVe?.
If you always want to transfer complete waveforms, set DATa:STARt
to 1 and DATa:STOP to 2500.
Examples
Related Commands
Group
Syntax
Arguments
Summary of Contents for TDS200 Series
Page 4: ......
Page 12: ...Preface viii TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 13: ...Getting Started ...
Page 14: ......
Page 16: ...Getting Started 1 2 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 17: ...Syntax and Commands ...
Page 18: ......
Page 32: ...Command Syntax 2 14 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 52: ...Command Groups 2 34 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 196: ...Command Descriptions 2 178 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 197: ...Status and Events ...
Page 198: ......
Page 223: ...Programming Example ...
Page 224: ......
Page 227: ...Appendices ...
Page 228: ......
Page 236: ...Appendix B Factory Setup B 6 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 237: ...Glossary and Index ...
Page 238: ......