Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−1 11
MATH:FFT:VERtical:SCAle
NOTE
. You can use this command with a TDS1000 or TDS2000
series, as well as a TDS200 series with a TDS2MM measurement
module.
Sets or queries the FFT math waveform vertical zoom factor.
Vertical
MATH:FFT:VERtical:SCAle <NR3>
MATH:FFT:VERtical:SCAle?
<NR3>
specifies the vertical zoom factor, where the axis of
magnification is the center horizontal graticule line. Valid zoom
factors are 0.5, 1, 2, 5, and 10. If other values are entered the
oscilloscope rounds the value of <NR3> to the nearest valid zoom
factor.
MATH:FFT:VERtical:SCAle 5
sets the FFT waveform vertical zoom factor to X5.
MATH:FFT:VERtical:SCAle?
might return
2.0E0
, indicating that the zoom factor setting is X2.
MEASUrement? (Query Only)
Returns all oscilloscope measurement parameters.
Measurement
Group
Syntax
Arguments
Examples
Group
Summary of Contents for TDS200 Series
Page 4: ......
Page 12: ...Preface viii TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 13: ...Getting Started ...
Page 14: ......
Page 16: ...Getting Started 1 2 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 17: ...Syntax and Commands ...
Page 18: ......
Page 32: ...Command Syntax 2 14 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 52: ...Command Groups 2 34 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 196: ...Command Descriptions 2 178 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 197: ...Status and Events ...
Page 198: ......
Page 223: ...Programming Example ...
Page 224: ......
Page 227: ...Appendices ...
Page 228: ......
Page 236: ...Appendix B Factory Setup B 6 TDS200 1000 2000 Series Oscilloscope Programmer Manual ...
Page 237: ...Glossary and Index ...
Page 238: ......