Performance Veri
fi
cation
5.
Connect the probe to any channel (1–4)
6.
Degauss the probe.
7.
Clamp the current probe around the HF
current loop. Verify that the arrow indicator
on the probe points away from the pulse
generator.
8.
The rise time is displayed on the
oscilloscope. Record the rise time
measurement in the Test Record.
9.
Calculate the probe bandwidth using the
measured rise time in the following formula:
10.
Record the calculated bandwidth value in
the test record.
11.
Disable the generator output.
End of procedures.
70
20 A Current Probes Instruction Manual