Speci
fi
cations
Typical Characteristics
Typical characteristics describe typical but not guaranteed performance.
Table 2: Typical electrical characteristics
Characteristic
Description
Maximum continuous current – DC and low frequency
20 A RMS
Maximum peak current
TCP0020 with 50
Ω
oscilloscope termination
50 A maximum peak pulse (pulse width
≤
20
μ
s)
TCP0020 & TCP2020 with 1 M
Ω
oscilloscope termination
100 A maximum peak pulse (pulse width
≤
10
μ
s)
Displayed RMS noise
≤
1.0 mA RMS (limit measurement bandwidth to 20 MHz)
Insertion impedance
Signal delay
~14.5 ns
Maximum voltage on bare wire
150 V CAT II
Maximum voltage on insulated wire
300 V CAT II
Maximum Amp·Second product
1000 A·
μ
54