Performance Veri
fi
cation
Equipment Required
Table 6 lists the equipment required for the performance veri
fi
cation procedure.
Table 6: Test equipment
Description and quantity
Performance requirement
Recommended example
1
Oscilloscope
TekVPI interface, 500 MHz or greater bandwidth
Tektronix DPO4000
High Amplitude Pulse
Generator
Risetime <500 ps, pulse width
≥
100 ns, amplitude
>10 Vpp into 50
Ω
Picosecond Labs 2600C
Calibrator
DCV: 0.2% accuracy, 0 to ±1.5 V, square wave output
ACA: 0.25% accuracy, 0 to ±6 A, square wave output
Fluke 9100
TekVPI Calibration/Veri
fi
cation
adapter
2
TekVPI interface
067-1701-xx
Digital multimeter (DMM)
≥
2 volt range,
≥
1% accuracy
Keithley 2700
DC current loop
5 turns 18 AWG coated wire on 3 inch form
See instructions that follow
HF current loop
50
Ω
±0.5%, BNC male connector
067-2396-xx
Adapter
BNC-to-dual banana
103-0090-xx
Adapter
2
SMA male-to-BNC female
015-1018-xx
BNC cable
50
Ω
, 0.76 m (30 in) length
012-0117-xx
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers
2
Required for TCP0020 probe only
62