Maintenance
5- 8
SDA 601 Serial Digital Analyzer
d.
Factory Reset
see page 3--6
5.
Flash Diags.
a.
Checksum Memory — Press
ENTER
to calculate a checksum of all 512
Kbytes of flash memory. The calculation will require approximately 7
seconds.
Typically, you will use the next three flash diagnostic items (
b
,
c
, and
d
)
together to confirm that the instrument can erase, read, and write to flash
memory. A useful approach is to:
H
Read byte 1000 (
b
);
H
erase sector 1, which begins at byte 10000 (
d
);
H
read byte 10000 (
b
) to confirm the erase (the byte value should be FF);
H
then write a new value into the byte (
c
); and finally
H
read it back (
b
), to confirm the write.
b.
Read Byte 10000 — Press
ENTER
to read the content of byte 10000.
c.
Write Byte 10000 — Enter a number from 0 to 255 with the number
keys (see Figure 5--2), then press
ENTER
.
d.
Erase Sec 10000 — Press
ENTER
to erase sector 1, which begins with
byte 10000.
e.
Read SLM Cals — Developed for use during design and manufacture;
not intended for use in the field.
f.
Write Byte Loop — Developed for use during design and manufacture;
not intended for servicing the instrument. Pressing
ENTER
will put the
SDA601 into an infinite loop; switch the instrument Off to exit the loop.
6.
Calibration.
a.
Calibrate SLM
see page 4--2
b.
Adj Serial Pot
see page 4--4
c.
Show Curr Temp — Press
ENTER
to check the internal instrument
temperature in “A/D units,”
not
degrees F or C.
d.
Set Manuf Watch — Developed for use during manufacture; not
intended for servicing the instrument. Pressing
ENTER
will initiate
special, undocumented Watch-mode “conditions.”
Summary of Contents for SDA601
Page 4: ......
Page 10: ......
Page 15: ...Specifications...
Page 16: ......
Page 21: ...Operating Information...
Page 22: ......
Page 29: ...Operating Information SDA 601 Serial Digital Analyzer 2 7 Figure 2 5 A Printed Analysis Report...
Page 49: ...Performance Verification...
Page 50: ......
Page 69: ...Adjustment Procedures...
Page 70: ......
Page 78: ...Adjustment Procedures 4 8 SDA 601 Serial Digital Analyzer...
Page 79: ...Maintenance...
Page 80: ......
Page 93: ...Replaceable Electrical Parts...
Page 94: ......
Page 104: ......
Page 105: ...Diagrams...
Page 106: ......
Page 108: ...Diagrams and Circuit Board Illustrations 7 2 SDA 601 Serial Digital Analyzer Data Path Diagram...
Page 111: ......
Page 122: ......
Page 123: ...Replaceable Mechanical Parts...
Page 124: ......
Page 129: ...Index...
Page 130: ......