Performance Verification Procedures
3- 12
SDA 601 Serial Digital Analyzer
d.
Set the cable simulator to 25 meters (if you are using a variable
attenuator, set it to 2.5 dB).
e.
Verify
that 1 or 2 more meter segments are “on” than in step
c
.
10.
Current Time
a.
Press the SDA601
Time (F)
key.
b. Verify
that the current date and time are correct.
I/O Menu.
11.
Highlighting
a.
On the TSG601, press
F
to select the “Limit Ramp” signal. Confirm that
the limit ramp signal is visible on the left display of the video measure-
ment set.
b.
Enter the SDA601 Watch menu (press
Shift
, then
Watch
). The LCD will
resemble Figure 3--10.
Figure 3- 10: The First Watch Menu Display
c.
Press the SDA601
"
key once to toggle the second line to “Watch.”
Press
Enter
to confirm the change.
d.
Press
Watch
.
e.
If the OSD is present on the left display of the video measurement set,
press
Insert On/Off
(on the SDA601) to toggle the OSD off.
f.
Enter the SDA601 I/O menu (press
Shift
, then
Insert On/Off
). The LCD
will resemble Figure 3--11.
Figure 3- 11: The I/O Menu/Highlighting Item Display
Summary of Contents for SDA601
Page 4: ......
Page 10: ......
Page 15: ...Specifications...
Page 16: ......
Page 21: ...Operating Information...
Page 22: ......
Page 29: ...Operating Information SDA 601 Serial Digital Analyzer 2 7 Figure 2 5 A Printed Analysis Report...
Page 49: ...Performance Verification...
Page 50: ......
Page 69: ...Adjustment Procedures...
Page 70: ......
Page 78: ...Adjustment Procedures 4 8 SDA 601 Serial Digital Analyzer...
Page 79: ...Maintenance...
Page 80: ......
Page 93: ...Replaceable Electrical Parts...
Page 94: ......
Page 104: ......
Page 105: ...Diagrams...
Page 106: ......
Page 108: ...Diagrams and Circuit Board Illustrations 7 2 SDA 601 Serial Digital Analyzer Data Path Diagram...
Page 111: ......
Page 122: ......
Page 123: ...Replaceable Mechanical Parts...
Page 124: ......
Page 129: ...Index...
Page 130: ......